What is the Response Time of the 701260 Analog Voltage Module When in RMS Observation Mode?

Solution:

The measured output response time of the 701260 Analog Voltage Input Module, when observing RMS is:
  • Rising (0 to 90% of 10 div): 100ms
  • Falling (100 to 10% of 10 div): 250ms
An attached JPEG displays the results of the 701260 output response to a step-input with a 60 Hz sine wave. Two zoom windows are displayed, the bottom-left zoom windows is the step-down response and the right window is the step-up response. The response time can be measured only when observing RMS.  