Digital Insulation Tester MY40 (DISCONTINUED)

Notice: This product was discontinued on Jul 13, 2018. See this replacement product:

The MY40 insulation tester is used to measure the Insulation resistance that represents the state of insulation of electric equipment or circuits; It is one of the important measurement parameters in terms of safety and security. The MY40 comes with 4 voltage/resistance ratings on an easy-to-view large display, and incorporates a double-action safety mechanism to keep the user from harm.

Specifications - MY40 Digital Insulation Tester

Width 125 mm
Height 103 mm
Depth 53 mm
Display 3 1/2-digit LCD; 4000 count; backlight-illuminated; logarithmic bar graph; extension bar graph-no fluctuations, as the display shows the digits of a reading in the order in which each digit settles.
Comparator Function The MY40 alerts you by turning on the LOW symbol and sounding the buzzer if the measured value is smaller than the reference value. You can allocate as many as three user-defined reference values to each rating. The factory-set defaults are 0.1 MΩ, 0.2 MΩ and 0.4 MΩ.
Memory Function For each rating, you can save as many as 20 measurements at desired memory address numbers.
Automatic Discharge Function The MY40 automatically begins discharge when you turn off the MEAS switch. You can monitor the state of discharge by checking the bar graph and make sure discharge is complete by checking that the segment bar disappear.
High-Voltage Indicators The high-voltage symbol and LED lamp come on to alert you when the MY40 is in insulation testing mode or if any voltage remains to be discharged.
Live-Line Alarm If you apply an AC voltage of approximately 40 V or higher across the input terminals, the MY40 alerts you by blinking the LED lamp and sounding the buzzer.
Overrange Input Alarm If the voltage being measured exceeds 600 V during AC voltage measurement, the MY40 alerts you by flashing the Maximum Value indicator and sounding the buzzer.
Auto-Hold Function The tester retains the measured resistance for approximately 5 seconds after the MEAS switch is turned off.
Batteries Four AA (R6P) batteries
Weight 420 g
Ambient Temperature/Humidity Ranges 23 ±5 /45-75% RH
Standard Accessories Batteries (Qty - 4)
Earth Probe (98002) (Qty - 1)
Line Probe (98001) (Qty - 1)
Protection Cover (93013) (Qty - 1)
Shoulder Strap (99005) (Qty - 1)
User's manual (Qty - 1)

 

Tolerance Mentioned Conditions

Max. Deviation from Zero Scale Value 6 digits
Indication of mark on bar graph Approx. 4000 MΩ min. (500 V/1000 V)
Approx. 400 MΩ min. (125 V/250 V)
No-Load Voltage 130% max. of rated voltage
Rated Measuring Current 1 mA (0 to 20%) when in first effective measuring range
Max. Short-circuit Current 2 mA

 

Testing Performance

Model Rating Range Option Resolution Measuring Range Tolerance Lower Limit of measured Ω Rated Current Central Scale Value
MY40-01 125V/200MΩ 0.4000 0.1kΩ 0-0.0199MΩ ± (5%of rdg+6dgt) 0.125MΩ 1mA 5MΩ
4.000 1kΩ 0.0200-20.00MΩ* ± (2%of rdg+6dgt)
40.00 10kΩ 20.01-200.0MΩ ± 5%of rdg
200.0 100kΩ    
250V/200MΩ 0.4000 0.1kΩ 0-0.0499MΩ ±(5%of rdg+6dgt) 0.25MΩ 1mA 5MΩ
4.000 1kΩ 0.0500-20.00MΩ* ± (2%of rdg+6dgt)
40.00 10kΩ 20.01-200.0MΩ ±5%of rdg
200.0 100kΩ    
500V/2000MΩ 4.000 1kΩ 0-0.999MΩ ± (5%of rdg+6dgt) 0.5MΩ 1mA 50MΩ
40.00 10kΩ 1.000-500MΩ* ± (2%of rdg+6dgt)
400.0 100kΩ 501-2000MΩ ± 5%of rdg
2000 1MΩ    
1000V/2000MΩ 4.000 1kΩ 0-1.999MΩ ± (5%of rdg+6dgt) 2MΩ 0.5mA 50MΩ
40.00 10kΩ 2.000-1000MΩ* ± (2%of rdg+6dgt)
400.0 100kΩ 1001-2000MΩ ± 5%of rdg
2000 1MΩ    

Looking for more information on our people, technology and solutions?


Contact Us

Precision Making

Top