How do I Perform a Shunt Calibration With the ScopeCorder Strain Gage Module?
To perform a shunt calibration using the 701271 strain gage module and the ScopeCorder, please complete the following procedures: Connect the strain module 701271, bridge head 701957/701958 and the strain gauge Attach...
item: 51 | http://tmi.yokogawa.com/vn/service-support/faqs/51/
Why Does the Self Test Fail on the WE7111 Digital Oscilloscope Module?
You must use a probe with a attenuator of "10:1", otherwise the self test will not run correctly. Also, before executing the self-test, please allow for a 30 minutes warm-up time. If the instrument does not warm up, the ...
item: 1175 | http://tmi.yokogawa.com/vn/service-support/faqs/1175/
Can I use a 100Base-TX Ethernet Connection With the WE7000 Series?
Yes, you can use a 100Base-TX/10Base-T with the WE7052 Fast Ethernet Module....
item: 1153 | http://tmi.yokogawa.com/vn/service-support/faqs/1153/
What is the Difference Between the WE7281 and the WE7282 4-Ch 100 kS/s D/A Modules?
The output connectors are different. The WE7281 has a clamp type terminal, and the WE7282 has a BNC terminal. Also, the maximum common mode voltage, and maximum voltage between channels etc. vary in their specifications ...
item: 1152 | http://tmi.yokogawa.com/vn/service-support/faqs/1152/
Why Doesn't the Math Waveform Display in Real Time When in Free Run Mode?
In free run mode, the execution timing is different depending on the type of computation. In the case of computations such as FFT, Filter, pulse width, etc. various data are arranged in a chronological direction. Calcu...
item: 1177 | http://tmi.yokogawa.com/vn/service-support/faqs/1177/
How can I Express an Arbitrary Exponentiation Other Than the Square and Cube Computation Function in XViewer?
You can express any arbitrary exponentiation, in Xviewer or DL series instruments, by using the LOG (common logarithm) and EXP (exponential) functions. For example, for C1 raised to the exponent of 0.2, the math functio...
item: 575 | http://tmi.yokogawa.com/vn/service-support/faqs/575/
When Adjusting a WE Module With the Adjustment Software, Which Slot do you use to Install the Module?
When using the Adjustment Software to adjust a WE7000 input module, the module should be installed in slot 2 of the main station. The slot labels on the WE Main Station begin with "0". Therefore, install the module in sl...
item: 1037 | http://tmi.yokogawa.com/vn/service-support/faqs/1037/
WE7000 Series Module Become Unresponsive When the Main Power Switch of the Measuring Station is Turned OFF and Then Immediately ON Again
You must wait at least five seconds after turning the power switch to the measuring instrument OFF before turning it back ON again. If you wait less than five seconds, there is a possibility for a malfunction to occur in...
item: 1172 | http://tmi.yokogawa.com/vn/service-support/faqs/1172/
Is There an Input Filter on the Digitizer Modules?
The WE7275 2-CH, 1 MS/s isolated digitizer module comes with an up-to-40 kHz anti-aliasing filter. Also, each digitizer module (WE7275/WE7271/WE7272/WE7251/WE7245) includes a low-pass filter. The cut-off frequency varies...
item: 1146 | http://tmi.yokogawa.com/vn/service-support/faqs/1146/
What is the Update Cycle of the D/A Output of the WE7141 100Mhz Universal Counter Module?
The update cycle changes depending on the acquisition mode and the gate time during frequency measurement. When the acquisition mode is One Shot, the update cycle is approximately 0.5-1 seconds. However for frequency mea...
item: 1149 | http://tmi.yokogawa.com/vn/service-support/faqs/1149/