What is the Maximum Visible Light Source Distance That can be Seen by the AQ7275?
The maximum visible light source distance that can be seen by the AQ7275, through a standard SM fiber, is approximately 5 km or 3.1 miles. ...
item: 518 | http://tmi.yokogawa.com/vn/service-support/faqs/518/
Is There a Limitation on the Number of Characters That can be Used on the DL1540?
Up to 25 characters can be entered and displayed. If you attempt to edit a preexisting comment, only the first nine characters are displayed. In this case, you must reenter characters after the ninth character. ...
item: 1215 | http://tmi.yokogawa.com/vn/service-support/faqs/1215/
How do I set a Gage Factor Greater Than 2.2 for the 701271 Strain Input Module?
The Gage Factor setting for the 701271 Strain Input Module can be directly inputted within the range of 1.90 to 2.20. However, it is possible to indirectly set the Gage Factor to a larger value by setting the constant A ...
item: 569 | http://tmi.yokogawa.com/vn/service-support/faqs/569/
Why are the Input Terminals on the Rear Panel of the Power Meter?
The input terminals on all Yokogawa power meteras are located on the rear panel. This takes into account safety when handling the measuring instrument. The signal input to the power meter normally carries high voltage...
item: 1269 | http://tmi.yokogawa.com/vn/service-support/faqs/1269/
How do I Upgrade the Firmware for the AQ2201/AQ2202 Frame Controller?
Please download and refer to the attached document for complete instructions on how to upgrade the firmware for the AQ2201/AQ2202 Frame Controller. ...
item: 300 | http://tmi.yokogawa.com/vn/service-support/faqs/300/
How Much Residual Noise Should There be on a Yokogawa Oscilloscope?
The true noise floor of a Yokogawa oscilloscope is ≤ 2-3mVpp. The DL9000, for example, has approximately 280uV residual noise at 2mV/Div with input termination set to 50Ω. The residual noise may be measured by making a ...
item: 531 | http://tmi.yokogawa.com/vn/service-support/faqs/531/
How do I Read the Measurement Values From the CSV File Saved by a WT1800 Using MATLAB?
A quick and simple way of accessing the measurement data from a CSV file saved by the WT1800 is to use the MATLAB import wizard. In MATLAB, simply click on File > Import Data.... You can also use the MATLAB built-in func...
item: 1278 | http://tmi.yokogawa.com/vn/service-support/faqs/1278/
What is the Step Response of the 701280 Frequency Input Module for a Sine Wave Filtered with an IIR LP Filter?
For the 701280 Frequency Input Module with a 4th order IIR 6Hz lowpass filter, a 60Hz sine wave (FM modulated with a square wave) was measured with deviations of ± 3Hz. Please download and view the attached image for a ...
item: 986 | http://tmi.yokogawa.com/vn/service-support/faqs/986/
Can I Upgrade the DLM2000 Firmware Using USB Flash Drive?
If your DLM2000 has firmware version 1.245 or later, then you can use a USB flash drive to upgrade the firmware of your unit. Note:DLM2000 refers to models DLM2024, DLM2022, DLM2034, DLM2032, DLM2052, DLM2054...
item: 523 | http://tmi.yokogawa.com/vn/service-support/faqs/523/
What are the RS232 Cable Pinouts Used by the WT & PZ Series Power Analyzers?
Please refer to the attached Excel spreadsheet for complete information regarding the RS232 cable pinouts used with by the WT & PZ Series Power Analyzers. ...
item: 481 | http://tmi.yokogawa.com/vn/service-support/faqs/481/