Why is the Display Update Rate Slower Than the Data Update Rate Setting on the WT500?

The actual display update rate for the WT500 will depend on the input signal and the trigger setting. In addition, there may be a very small trigger delay (several milliseconds) if the input signal does not match the trigger condition. For example, when the update rate is set to 1 sec and the trigger mode is set to Auto, if the input signal does not match the trigger condition then the waveform measurement is performed after the sum of 1 sec and the timeout period. The timeout period is 100ms.
 
The display update interval, specified by the WT500 instruction manual, is listed as:
 
Numeric Display
  • When the data update interval is 100 ms and the numeric display is set to 4, 8, or 16 items, the display update interval is 200 ms.
  • When the data update interval is 200 ms and the numeric display is set to Matrix, ALL Items, Single List, or Dual List, the display update interval is 500 ms.
  • When the data interval is 200 ms and the numeric display is set to Matrix, ALL Items, Single List, or Dual List, the display update interval is 400 ms.
Waveform Display
  • When the data update interval is within the range of 100 ms to 1 s, the waveform display and data update intervals are approximately 1 s.
  • When the data update interval is 1 s or longer, the waveform display and data update rate intervals are the same as the data update interval but may be longer depending on the trigger settings.
Other Display
  • When the data update interval is within the range of 100 ms to 500 ms, the data and display update interval of the trend, bar graph, and vector display is 1 s. When the data update interval is set to 1 s or longer, the various display and data update intervals are the same as the data update interval.
Response Time
  • The maximum response time is the data update interval x 2

 

Related Products & Solutions

WT500 Mid-Range Power Analyzer

The WT500 Power Analyzer excels at single-phase and three-phase power measurements. Standard features include a color TFT display and USB interface for communications and memory. The instrument has a basic power accuracy of 0.1%, maximum inputs of 1000 V, 40 A and a measurement bandwidth of DC to 100 kHz.