PSD Switching Measurement with High Definition Oscilloscope

Download (216 KB)

PSD*Switching Measurement with High Definition Oscilloscope

* Power Semiconductor Device

Issue:

The switching loss ("Power" in the example) is calculated by multiplying the output voltage(VDS) and current(Id) in the switching period (turn ON/turn OFF). The ON voltage is only a few volts for an output voltage of several hundred volts, but with a typical 8-bit oscilloscope, this is only 1 to 2 bits of AD resolution, and highly accurate measurements are not possible.

Solution:

The DLM5000HD with 12-bit ADC has 16 times the resolution of conventional 8-bit oscilloscopes, making it ideal for such wide dynamic range measurements. In addition, next-generation devices such as SiC/GaN have a high-speed rising switching voltage, which causes high-voltage ringing, and suppressing it is an important point in designing inverters.  High Definition Oscilloscope is also effective for such ringing/noise observations.

Example: SiC MOSFET double pulse test waveform

Example: SiC MOSFET double pulse test waveform

Related Industries

Related Products & Solutions

701927 Differential Probe 1400V / 150 MHz

  • 701927 Isolated Differential Probe
  • 150 MHz 
  • Max. Differential Voltage: ±1400 V (DC + ACpeak) at 500:1
  • Probe Power: dedicated probe interface For DLM, DL, and SB5000 series

701928 Current Probe 100 MHz / 30 ARMS

701928 current probe

Front-end powered

Recognized automatically

Adjust zero position from operation of DLM2000, DLM3000, DLM4000, DL6000/DLM6000, DL9000/DLM9000, SB5000

701978 Differential Probe 1500 V / 150 MHz

  • 701978 Differential Probe for Oscilloscope 
  • 1500 V / 150 MHz
  • 500:1 / 50:1
  • Max. Differential Voltageļ¼š±1500 V (DC + ACpeak)

702916 Current Probe 120 MHz/ 0.5 ARMS, 5 ARMS, 30 ARMS

Frequency bandwidth: DC to 120 MHz
Maximum continuous input range: 0.5 Arms/5 Arms/30 Arms
Functions with Digital Oscilloscopes, ScopeCorders and other waveform measuring instruments

Precision Making

Top