700929 ISOLATED PASSIVE PROBE 1000V / 100 MHZ

Passive Probe, 1000Vp CAT2, 100MHz, 10:1, 10MΩ, 1.5m, Isolated Probe, Safety Iso-Probe.  For use with 701250, 701251, 701260, 701280, 701281, 720210 isolated-input modules ONLY.  For use with DL850 series, DL750/DL750P, SL1400, and SL1000.  (This is the "new style" 700929 Iso-Probe for ScopeCorder.)

See also:

New 700929 Adapters
701947 100:1 Isolated Probe (3,540Vpeak CAT I, 1000Vpeak Cat II)
M1223WN-A Push-On Large-Jaw Safety Clip (high-side)
M1223WP-A Push-On Male Banana Plug (high-side)
M1223WQ-A Reference Lead & Adaptor (low-side)
M1223WR-A Large Jaw Safety Reference Clip

Get a Quote
Ask an Expert

 
 

Specifications

  • Bandwidth: DC to 100MHz (-3dB)
  • Attenuation ratio: 10:1
  • Connector Type: BNC
  • Input Resistance: 10MΩ
  • Input Capacitance: 18pF
  • Compensation Range:
  • Rise Time: 1.4ns
  • Maximum input voltage: 1,000V (DC + ACpeak) (subject to specification of the module), derate above 500kHz.
  • Common-Mode Input Voltage: Limited to the specifcatin of the Module used with this probe.
  • Cable length: 1.5m
  • Standard accessories included: B9940WX Safety Plug-On Clip (pincher), and Safety Clip Lead (B9940WX)
  • Safety Rating (Probe Alone): EN61010-031:2002, CAT III