How to Select Current Probe Attentuation on DLM2000 Mixed Signal Oscilloscope Via Communication?

When using voltage probes, please set the channel probe mode using the command :CHANnel<x>:PROBe 10.00 or 0.01.
When using  current probe, please set the channel probe mode using the command :CHANnel<x>:PROBe C10 or C0_01.
 
:CHANnel<x>:PROBe[:MODE]
     
Funciton. It sets or queries a probe's attenuation rate of each channel.
Construction. :CHANnel<x>:PROBe[:MODE] {<NRf>|C0_001|C0_002|C0_005|C0_01|C0_02|C0_05|C0_1|C0_2|C0_5|C1|C2|C5|C10|C20|C50|C100|C200|C500|C1000|C2000}
:CHANnel<x>:PROBe[:MODE]?
 <x> =from 1 to 4 (1 or 2 for DLM2022, DLM2032, and DLM2052)

Related Products & Solutions

DLM2000 Mixed Signal Oscilloscopes

200, 350, and 500MHz mixed-signal oscilloscopes for every engineer. Best-in-class performance in usability, acquisition, analysis, and display—all at a price you can digest. Options include serial bus, vehicle bus, and power supply analysis functions.

Mixed Signal Oscilloscopes

  • Yokogawa mixed signal oscilloscopes
  • Simultaneous, time correlated observations
  • Analysis of analog with digital (logic) signals
  • Troubleshooting electrical anomalies, measuring parametric values, monitoring signals

Oscilloscopes

  • Digital oscilloscopes
  • High-speed sampling
  • Range of bandwidths for electronic device design and development
  • Advantages of oscilloscope and multi-channel data recorder

Precision Making