Why Won't my USB Flash Drive Read/Write on the DLM2000?

Please verify the following items: USB does not exceed 16 GB Latest firmware is installed on DLM2000 USB Format is FAT32 USB works properly on PC Test USB on another USB port If the problem persists, please contact Yok...

item: 1293  |  http://tmi.yokogawa.com/de/service-support/faqs/1293/

How do I Clear the Menu Screen and Message on the DL1540 from the PC?

The menu screen and messages are cleared when commands such as "start" and "cursor on" are issued. ...

item: 1203  |  http://tmi.yokogawa.com/de/service-support/faqs/1203/

Is the Delay Measurement Value a Fixed-Point or Floating Point Number?

The Delay Measurement value is a Fixed-Point Number.The Delay Measurement feature on the DL850 comes in two modes, Time and Degree. The output of the Delay Measurement feature is a time-based value and not an amplitude ...

item: 1001  |  http://tmi.yokogawa.com/de/service-support/faqs/1001/

How do I Programmatically Read Average Active Power Right Before the End of the Integration Interval While in Continuous Integration for the WT230?

You can read the average active power during continuous integration mode (just before the integral resets) for the WT230, by monitoring the status of the ITG or ITM bits of the extended event register.   Bit 1 ITG is u...

item: 982  |  http://tmi.yokogawa.com/de/service-support/faqs/982/

Can Xviewer Combine the Binary Format Data Files Saved by Different Products?

Yes, Xviewer has the capability of opening multiple files from different DL, SL, and WE series instruments into a single window display. The only requirement for viewing files from different products is that the sample ...

item: 539  |  http://tmi.yokogawa.com/de/service-support/faqs/539/

What is the Maximum Input Power for Pulsed Light Measurements on the AQ6370B?

The maximum input power that the AQ6370B can measure for pulsed light measurements is +20dBm per resolution at full span, per channel. Keep in mind that this limit is applicable to an averaged power of pulsed light, not ...

item: 161  |  http://tmi.yokogawa.com/de/service-support/faqs/161/

Why Won't Xviewer Open the CSV ASCII File Created by the ScopeCorder?

To open the CSV ASCII file saved by your ScopeCorder: Start Xviewer Open a New Viewer Window Drag and drop your CSV file into the empty viewer window Additional Note:ScopeCorder refers to the DL750/DL750P, DL850/DL850V,...

item: 56  |  http://tmi.yokogawa.com/de/service-support/faqs/56/

INTEG1 vs INTEG2 - as demonstrated on DL850 G3

INTEG1 vs INTEG2 - as demonstrated on DL850 G3   Examples of two different ways to INTEGRATE on the DL850/G3 DSP.   One attachment. ...

item: 1311  |  http://tmi.yokogawa.com/de/service-support/faqs/1311/

Can the DL1540C Load the Binary Data File of the DL1540?

No, the DL1540C cannot load the binary data file of the DL1500. The DL1540C can only load the binary data file saved by the DL1540C. The DL1540CL can load the binary data file saved by the DL1540C and DL1540CL...

item: 617  |  http://tmi.yokogawa.com/de/service-support/faqs/617/

Is There a Way to Recover a DL850 WDF File That Cannot be Opened in Xviewer?

If the DL850 WDF file was saved through the hard disk recording feature (HD Recording), then the file may be recovered by using the Finalize feature in Xviewer File Utility. Files that were saved through the normal save ...

item: 1284  |  http://tmi.yokogawa.com/de/service-support/faqs/1284/