This high-voltage differential probe features a wide frequency range from DC to 400 MHz, excellent frequency response flatness, and superior common mode rejection ratio (CMRR) characteristics.
Bandwidth |
: |
DC to 400 MHz |
Attenuation |
: |
1000:1/100:1 |
Max. Input Voltage (to Ground) |
: |
±2000V (DC + ACpeak) |
Max. Differential Voltage |
: |
±2000V (DC + ACpeak) at 1000:1 Attenuation |
Max. Non-Destructive Input Voltage |
: |
±3000V (DC + ACpeak) (within 5 seconds) |
Probe Power |
: |
Dedicated probe interface |
For use with DLM3000, DLM3000HD, DLM5000, DLM5000HD series |
Frequency Band | DC to 400 MHz (-3 dB) |
---|---|
Input Attenuation Ratio | 100:1 and 1000:1, User Selectable |
Input Impedance (to Ground) | Approximately 6 MΩ, Approximately 5 pF |
Maximum Allowed Differential Voltage | ± 2000V (DC+ACpeak) (1000:1) ± 200V (DC+ACpeak) (100:1) |
Maximum Common Mode Input Voltage (to Ground) | ± 2000V (DC+ACpeak) |
Maximum Non-Destructive Input Voltage (to Ground) | ±3000V (DC + ACpeak) (within 5 seconds) |
Common Mode Rejection Ratio (CMRR) (Typical) | -80 dB (60 Hz), -50 dB (10 MHz), -30 dB (100 MHz), -20 dB (400 MHz) |
Connector Type (Power Supply) | Yokogawa Probe Interface Only (Applicable to DLM3000, DLM3000HD, DLM5000, and DLM5000HD) |
External Dimensions (Approximate) | Head : 173.5(H) x 58(W) x 26.5(D) mm Connector : 80(H) x 29(W) x 24(D) mm |
Total Length | Approximately 2.4 meters |
Weight | Approximately 300 grams |
Measurement Category | CAT II 1000 V, CAT III 600 V |