If There is a Temperature Coefficient Listed in the Specifications, When is the Coefficient to be Applied?

If there is a temperature coefficient listed in the specifications of the instrument's user manual, that coefficient is to be applied in the following conditions:
  1. The unit is calibrated within the Standard Operating Conditions and the temperature of the unit changes in the range of the Standard Operating Conditions after calibration
  2. The unit is calibrated within the Standard Operating Conditions and the temperature of the unit changes to outside the range of the Standard Operating Conditions after calibration
  3. The unit is calibrated outside the Standard Operating Conditions and the temperature of the unit does not change after calibration
  4. The unit is calibrated outside the Standard Operating Conditions and the temperature of the unit changes after calibration

Case1:
 

If the unit is calibrated in the Standard Operating Conditions and if the temperature of the unit does not change, the accuracies of the measurement are only fundamental accuracies. If the unit is calibrated in the Standard Operating Conditions but if the temperature of the unit is changed in the range of the Standard Operating Conditions after the calibration, the temperature coefficients (only Zero) will be applied to the fundamental accuracies.

For example:
If the 701250 is calibrated at 23°C and the temperature does not changed after the calibration, the measurement accuracy is 5 mV/div to 20 V/div: ±(0.5% of 10 div).
If the 701250  is calibrated at 23°C, and the temperature changes to 25°C (in the Standard Operating Conditions), the measurement accuracy is 5 mV/div to 20 V/div: ±(0.5% of 10 div)  and  Zero point: 5 mV/div to 20 V/div: ±(0.05% of 10 div)*(25°C-23°C).

 
Case2:
 
If the unit is calibrated within the Standard Operating Conditions and the temperature of the unit changes to outside the range of the Standard Operating Conditions after calibration, the temperature coefficients (Zero and Gain) is applied to the fundamental accuracies. The calibrated temperature value of the Zero coefficients and the calibrated temperature value of the Gain coefficients are not same.
Zero coefficients: the calibrated temperature value is the difference from the calibrated temperature. 
Gain coefficients: the calibrated temperature value is the difference from the Standard Operating Conditions.
       
For example:
If the 701250 is calibrated at 23°C, and the temperature changes to 30°C (outside the Standard Operating Conditions range), the accuracies are 5 mV/div to 20 V/div: ±(0.5% of 10 div) and Zero point: 5 mV/div to 20 V/div: ±(0.05% of 10 div)*(30°C-23°C) and Gain: ±(0.02% of 10 div)*(30°C-28°C). The Standard Operating Conditions are from 18°C to 28°C. So the Gain coefficients are calculated from 28°C.
 
 
Case3:
 
If the unit is calibrated outside the Standard Operating Conditions and the temperature of the unit does not change after calibration, the temperature coefficients(only Gain) is applied to the fundamental accuracies.
        
For example:
The 701250 is calibrated at the 30°C, but the temperature does not change, the accuracies are 5 mV/div to 20 V/div: ±(0.5% of 10 div) and Gain: ±(0.02% of 10 div)*(30°C-28°C). The Standard Operating Conditions are from 18°C to 28°C. So the Gain coefficients are calculated from 28°C.
 
 
Case4:
 
If the unit is calibrated outside the Standard Operating Conditions and the temperature of the unit changes after calibration, then the temperature coefficients (Zero and Gain) is applied to the fundamental accuracies.
        
For example:
The 701250 is calibrated at the 30°C, but the temperature becomes to 32°C(out side of the Standard Operating Conditions), the accuracies are 5 mV/div to 20 V/div: ±(0.5% of 10 div)  and  Zero point: 5 mV/div to 20 V/div: ±(0.05% of 10 div)*(32°C-30°C) and Gain: ±(0.02% of 10 div)*(32°C-28°C). The Standard Operating Conditions are from 18°C to 28°C. So the Gain coefficients are calculated from 28°C, but the zero coefficients are calculated from the 30°C calibrated point.


The DL750, DL850, and 701280 modules do NOT list a Temperature coefficient in their specifications

相关产品和解决方案

数据采集设备

Yokogawa数据采集系统具备出众的灵活性和性能,能测量、显示、存储甚至促动各种物理和电气现象。

波形测量仪器

Yokogawa在业界首推示波记录仪的概念,将数字示波器和多通道数据记录仪的优点集于一身。横河示波记录仪作为世界一流的示波记录仪具有高速采样能力和大测量带宽的优点,可广泛应用于电子工程的设计和开发过程。

横河示波器既有面向低成本市场的4CH MSO示波器DLM3000系列,也有业界全新的8CH MSO示波器DLM5000系列,在由智能控制的消费电子和工业驱动等电路设计的众多领域中,成为全球工程师的首选工具。

示波记录仪

我们的示波记录仪是灵活、强大的多通道测试和测量解决方案,将高精度示波器和传统有纸图表记录仪的优点集于一身。Yokogawa的示波记录仪除了能在它的大显示屏上显示信号之外,还可以将信号保存到记录纸和/或存储介质中以便长期储存(如30天或更长),此外它还可以捕捉并分析非常快的瞬变信号。

高速数据采集器和数采仪

Yokogawa高速数据采集系统提供业界领先的隔离、位分辨率、采样速率和存储长度,带有独立的通道硬件和使用简单的软件。是示波记录仪中最领先的技术。

Precision Making

返回顶部
WeChat QR Code
微信扫一扫
获得更多专业服务