CA500 Series Multi-Function Process Calibrator

The CA500 and CA550 are high-performance, multi-function calibrators designed for use in multiple applications. Featuring useful functions for field work, these multi-function calibrators help you calibrate field instruments with improved source and measurement accuracy. 

Two models are available, offering two different levels of accuracy:

  • The CA500 achieves an accuracy of 0.015% (DC current), 0.020% (resistance), and 0.3˚C (RTD). 
  • The CA550 achieves an accuracy of 0.010% (DC current), 0.015% (resistance), and 0.1˚C (RTD).

Multi-function

  • Sources and measures DC voltage, DC current, RTD, TC, resistance, frequency and pulse signals
  • Corresponds to 17 types of TC standard (JIS/IEC/DIN/ASTM/GOST R)
  • Corresponds to 14 types of RTD standard (JIS/IEC/GOST R)

Multiple source patterns

Linear sweep function
Continuously Source from 0% to 100%

Step sweep function
Change output in a staircase (step) pattern by specifying the number of steps. (The number of steps can be set from 2 to 20)

Program sweep function
Users can set the desired output value (%) and number of steps. (10 steps for CA500, 20 steps for CA550)

Design

Design

 

Functions

Automatic input/output testing *CA550 only
(Program sweep)

Automatic input/output testing is possible by setting source values for each step for a calibration target. Calibration results such as generated value, measured value, error rate, date/time, and pass/fail are saved in CSV format in the CA550 main unit.

By connecting the CA550 to a PC using a standard USB cable, the instrument can be recognized as a mass-storage device for data to be transferred to the PC.

HART Communication Protocol *CA550 only

Firmware Ver2.03 or later (released)

The following items are supported by HART communication function:

  • LOOP TEST

  • TagNo.
  • PV value (including reading of PV %value, AO value, SV value, TV value, QV value)

Read

  • LRV (Lower limit of range)
  • URV (Upper limit of range)
  • Damping

Read and Write

  • Trimming of 4 mA
  • Trimming of 20 mA
  • PV Zero

Write

*HART communication that supports some of the universal commands and common Practice commands.
*BRAIN communication only supports acquisition of Tag No.

Easy-to-view Display

CA500 features a Reflective LCD, providing improved outdoor visibility. Main display (generated/measured values) and Sub display (%, mV, Ω, etc.) allow required information at a work site to be confirmed at a glance.

Easy-to-view Display

Wiring information display function

A wiring diagram is displayed according to the function selected. This function allows a user to perform wiring while referring to a wiring diagram and prevents mis-wiring.

Wiring information display function

Thermocouple generation using TC Mini Plug

Using a TC Mini Plug together with a compensating lead wire enables generation of thermal electromotive force without an external RJ sensor.*
* A compensating lead wire needs to be prepared by customer.

Thermocouple generation using TC Mini Plug

Easy-to-use key operation

0%/100% keys
The source can be easily switched between 0% and 100% of range. Users can also set a desired value.

UP/DOWN keys
The output is changed in preset steps by pressing UP or DOWN key.

Easy-to-use key operation

Operation key layout
Keys related to generation and measurement are arranged collectively to allow easy and intuitive operation.

Operation key layout

SQUARE ROOT output

For 4-20 mA, 1-5 V ranges, users can choose between LINEAR and SQUARE ROOT output.

SQUARE ROOT output

 

Applications

20 mA SIMULATE

The CA500 series can be used as a transmitter simulator to perform a loop test. It sinks the set current from an external voltage source of instrumentation equipment.

20 mA SIMULATE

Two-wire Transmitter Loop Check

DC mA signals can be measured by supplying power to the transmitter from a 24 V DC power supply. DC mA signal measurement and zero-point check can be performed with an accuracy of 0.01% of reading (0.015% of reading for CA500).Two-wire Transmitter Loop Check

 

Zero point adjustment of HART transmitter

CA550 supports HART communication (Universal command/Common practice command).
Reading of HART device information, writing of LRV/URV, and trimming of analog output are possible.

Zero point adjustment of HART transmitter

TC SIMULATE

The CA500 series corresponds to 17 types of TC for sourcing. It achieves the high basic accuracy of 0.5°C (typical of type K), two times better than the previous model.
Also, input/output testing is possible with a single CA500/CA550, as it can measure output signals.

TC SIMULATE

RTD SIMULATE

CA500/CA550 corresponds to 14 types of RTD for sourcing. It achieves a high basic accuracy of 0.1°C* (typical of type Pt100), which enables it to operate a highly reliable test.
Additionally, input and output testing of temperature transmitters is possible at the same time.
*Accuracy for CA550

RTD SIMULATE

Pulse SIMULATE

This calibrator performs measurement of pulse signal integration from a flowmeter and generates a pulse to a receiver, such as integrating counter or pulse converter. Integration time can be set from 1 min to 60 min.

Pulse SIMULATE

 

LCA Reduced energy, emission of CO2, NOX and SOX about 14% to 22% compared to the previous model.

PDFResults of Life Cycle Assessment

98020 Lead cable for source

One set of 1 red and 2 black cables
Length: Approx. 1.7 m

98035 Lead Cables

Alligator Clip Cable (Red x 3pcs, Black x 1pce 1 set/ 1.7m)

99045 Binding Post (Red Black)

Used for terminating (attaching) a single wire or test lead.

99046 Binding Post (Red Red)

Used for terminating (attaching) a single wire or test lead.

A1421WL USB Cable

Total length: 2 m
USB2.0 Hi-Speed

98064 Lead Cables

Alligator clip, CAT I, for controlsignal only (under 70 V)One set of 1 red and 1 black cablesLength: Approx. 1.7m

90080 RJ Sensor

RJ (Reference Junction)

98026 Grabber Clip

A set of separate red and black clips (for 2 m long wires).

SU2006A Soft carrying case

Soft Carrying Case

90040 TC Mini Plug Set 1

K (Yellow) /E (Violet) /J (Black) /T (Blue) /R,S (Green) /B,U (White) /G (Red/Green) /D (Red/White) /C (Red) /N (Orange)

90045 TC Mini Plug Set 2

K (Yellow) /E (Violet) /J (Black) /T (Blue)

90046 TC Mini Plug Set 3

K (yellow)/ E (violet)/ J (black)/ T (blue)/ R•S (green)/ B•U (white)/ G (red, green)/ N (orange)

93026 Carrying case

Lead cables for source/measurement, terminal adapter, 6 spare batteries, fuse, AC adapter and Instruction Manual can be stored.

Looking for more information on our people, technology and solutions?


Contact Us

Precision Making