Always observe instruction manual cautionary notes.
Follow maximum voltage level restrictions (up to 300V AC).
Keep transformer jaws fully closed during current measurement.
AC Current Measurement:
Set range switch
Press jaw open/close lever and close jaws over one conductor only. The measured current value is shown on display.
| Description | Leakage Current Mini Clamp-on |
|---|---|
| Method of detection | Mean Value |
| Display | LCD (Digital display: 1999 counts) |
| Response Time | Approx. 2 seconds |
| Range Switching | Manual-range |
| Data hold | On all range |
| Operating temperature and humidity | 0 - 40°C, 85% RH or less (no condensation) |
| Influence of external magnetic field | 10 mA or less in proximity to a 14.4 mm-dia conductor carrying 100A |
| Influence of conductor position | Within 5 dgt for 0 to 50A, or 2% for 50 to 200A (10 mm-dia conductor at inside the jaw) |
| Influence of Residual Current | 10 mA or less in proximity to a 10 mm-dia conductor carrying 50A |
| Safety Standard | Conforms EN61010-1, EN61010-2-032 |
| Circuit voltage | 300 Vrms or less |
| Withstanding voltage | 3.7 kV AC for one minute |
| Power Supply | LR-44 x 2(3V) SR-44 x 2 |
| Battery Life | Approx. 15 hours (continuous) |
| Consumed current | Approx. 5 mA |
| Auto power-off | Approx. 10 minutes |
| Diameter of measurable conductor | 24 mm at maximum |
| Dimensions | Approx. 60 (W) x 149 (H) x 26 (D) mm |
| Weight | Approx. 120 g |
| Accessories | Batteries, carrying case (93033) User's manual (IM CL320) |
| DC Current Range | - |
|---|---|
| DC Current Resolution | - |
| AC Current Range | 20/200mA,200A |
| AC Current Resolution | 0.01mA |
| AC Current Frequency characteristics | 40~400Hz |
| AC Current Method of detection | Mean Value |
| AC Voltage Measurement | - |
| DC Voltage Measurement | - |
| Continuity Check Measurement | - |
| Frequency Measurement | - |
| Temperature Measurement | - |
| Recorder output Function | - |
| Waveform monitor output Function | - |
| Data hold Function | O |
| Auto hold Function | - |
| Peak Hold Function | - |
| Maximum value memory Function | - |
| Mean value display Function | - |
| Filter Function | O |
| Range Hold Function | O |
| Auto ZERO Adjust Function | - |
| Auto power-off Function | O |
At 23°C±5°C,75%RH or less
Accuracy: ±( % rdg + dgt) = ±(% readout + value of least significant digit)
| Parameter | Reference | Accuracy | |
|---|---|---|---|
| WIDE(40~400Hz) | 50/60Hz | ||
| AC current | 20mA/200mA | 2.0+4 (50/60Hz) | 3.0+5 (50/60Hz) |
| 200A(0~100A) | 5.0+6 (40~400Hz) | ||
| 200A(100.1~200A) | 5.0+4 (50/60Hz) | 5.0+5 (50/60Hz) | |