The 700924 high voltage differential probe is an active voltage probe with high speed capabilities for measuring floating and high-speed differential signals in combination with single-ended input oscilloscopes. This differential voltage probe's attenuation ratio is switchable between 100:1 and 1000:1.
| 
			 Bandwidth  | 
			
			 :  | 
			
			 DC to 100 MHz  | 
		
| 
			 Attenuation  | 
			
			 :  | 
			
			 1000:1/100:1  | 
		
| 
			 Max. Input Voltage  | 
			
			 :  | 
			
			 ±1400 V (DC + ACpeak) or 1000 Vrms CAT III  | 
		
| 
			 Max. Differential Voltage  | 
			
			 :  | 
			
			 ±1400 V (DC + ACpeak) at 1000:1 attenuation  | 
		
| 
			 Probe Power  | 
			
			 :  | 
			
			 Internal battery or probe power supply  | 
		
| 
			 B9852MJ power cable is sold separately  | 
		||
Applicable for all Digital Oscilloscopes or ScopeCorders.
With increases in the speed of measured signals in recent years, it has become necessary to make floating measurements over a wide frequency band. The 700924 differential probe lets you make wide-band (100 MHz) differential input measurements by just connecting the probe to the input of a single-end input digital oscilloscope.
Oscilloscope input amplifiers are normally single-ended. They do not provide accurate measurements when used in combination with a 10:1 passive probe for measuring the gate-to-source voltage of a switching power supply, or potential difference with a common mode voltage in elements such as Insulated Gate Bipolar Transistors (IGBTs).
Furthermore, there is a risk that current will enter through the oscilloscope ground line and damage the oscilloscope.


| Input type | Balanced differential input | 
|---|---|
| Frequency band | DC up to 100 MHz (-3 dB) | 
| Input attenuation ratio | Switched between 1/100 and 1/1000 | 
| Input impedance | Approximately 4 MΩ, approximately 10 pF parallel | 
| Gain accuracy | ±2% (common mode voltage: 400 V or less) ±3% (common mode voltage: 1000 V or less)  | 
		
| Maximum allowed differential voltage  | 
			±350 V (DC + ACpeak) or 250 Vrms (1/100 range) ±1400 V (DC + ACpeak) or 1000 Vrms (1/1000 range)  | 
		
| Maximum common mode input voltage  | 
			±1400 V (DC + ACpeak) or 1000 Vrms (both 1/100 and 1/1000 ranges) | 
| Maximum input voltage | ±1400 V (DC + ACpeak) or 1000 Vrms (1/1000 range) | 
| Common Mode Rejection Ratio (CMRR) | -80 dB (60 Hz), -50 dB (1 MHz) | 
| Output voltage | ±3.5 V | 
| Output impedance | Used with 1 MΩ input impedance | 
| Power supplies | Dry cells, four R6P (SUM-3); when using the AC/DC adapter, use an output voltage of 6 V/200 mA or greater/center plus | 
| Dry cell life | Approximately 2 hours under continuous use | 
| External dimensions (approximate) | 207 mm x 83 mm x 38 mm¹ | 
| Total length | Approximately 1.5 meters | 
| Weight | Approximately 800 grams (excluding dry cells) | 
| Approved standards | EMC: EN55011: 1991 + A1: 1997 + A2: 1996 EN50082-1: 1992 Safety: EN61010-1: 1993 + A2: 1995 EN61010-2-031: 1994  | 
		
¹: 38 mm is listed as the maximum value.
Note: Dry cells and AC/DC adapter not included.
| Model Name | Code | Description | 
|---|---|---|
| Differential probe | 700924 | DC to 100 MHz band | 
Each 700924 differential probe comes standard with red and black pincher tips (one each).