How to Display Averaged Power Value During Integration on the WT1600 Digital Power Analyzer?

When the WT1600 Digital Power Analyzer is set into Integration mode, the averaged power (watt) values can be calculated and displayed. This is available only by using the User-Defined Function feature found in the MEASURE button menu.
 
The following equation must be entered in any one of the twenty Expression field:
WH(E1)/TI(E1)*3600
 
To display the calculated result, simply select and display the User Defined Function labeled F1 to F2 in the ITEM button menu.
 
The averaged power equation is pre-set on the WT500 and WT3000.

Related Products & Solutions

Precision Power Analyzer WT3000

With 0.02% accuracy and 1MHz bandwidth, the WT3000 delivers where the highest precision measurements are required. It is the industry standard for R&D work on inverters, motor drives, lighting systems and electronic ballasts, UPS systems, aircraft power, transformer testing, and other power conversion devices.