701938 Passive Probe 600V / 200 MHz

Bandwidth

:

DC to 200 MHz

Attenuation

:

10:1

Input Resistance

:

10 MΩ

Input Capacitance

:

13.5 pF

Max. Input Voltage

:

600 V (DC + ACpeak) or 424 Vrms CAT II

Probe length

:

1.5 m

For use with DLM2000 (200 MHz model)

How to Use Passive Oscilloscope Probe 701938

Use adequate attachment suitable for the point to measure. Before using the probe, adjust its capacitance by tuning the trimmer.

Adjustment

1. Connect the probe connector to the input of the oscilloscope, and connect the tip of the probe to the CAL signal output terminal.

2. Change the T/Div and the V/Div to get the display shown below. And tune the trimmer to get the correct waveform.

Specifications

 
701937

701938

701939

701943

701944

701945

701946

701949

702906

702907
Bandwidth 500 MHz 200 MHz 500 MHz 500 MHz 400 MHz 250 MHz 500 MHz 500 MHz 200 MHz 200 MHz
Max. input voltage 600 V (DC + ACpeak) 1000 Vrms 400 Vrms 1000 V (DC + ACpeak)
Attenuation ratio 10:1 100:1 10:1
Input impedance 10 MΩ
10.5 pF
10 MΩ
13.5 pF
10 MΩ
10.5 pF
10 MΩ
12.5 pF
50 MΩ
7.5 pF
10 MΩ
9.5 pF
10 MΩ
16 pF
10 MΩ
18 pF
Measurement category CAT II 600 V (DC + ACpeak) CAT II 1000 Vrms CAT II 300 Vrms CAT II 1000 V
(DC + ACpeak)
Compatibility DLM3000,
DLM3000HD,
DLM5000,
DLM5000HD
DLM2000 DL6000,
DLM2000,
DLM4000,
DLM6000
DL9000,
DL9500,
DL9700,
SB5000
DLM3000,
DLM3000HD,
DLM5000,
DLM5000HD
DL6000,
DLM2000,
DLM4000,
DLM6000
DLM3000,
DLM3000HD,
DLM5000,
DLM5000HD
DLM2000,
DLM4000
DLM3000,
DLM3000HD,
DLM5000,
DLM5000HD

Looking for more information on our people, technology and solutions?


Contact Us

Precision Making

Top