DL2700 Digital Oscilloscope (DESCONTINUADO)

Aviso: Este produto foi descontinuado em Veja este produto de substituição:

Visão Geral:

The performance varies depending on the firmware version. The details are shown as follows: Saving byte data to a 230 MB MO: 1 kW:4 s   10 kW:4 s   50 kW:6 s   100 kW:7 s 250 kW:11 s   500 kW:18 s   1 MW:32 ...

Visão Geral:

When saving measured data or set-up data, there is a limitation on the number of files that can be saved depending on the data format. The number of files that can be saved for the data format is listed below: Waveform ...

Visão Geral:

The data is saved in blocks of 8-bits and is expressed in decimal notation. In other words, if all 8-bits are zeroes, the decimal value is 0; if all 8-bits are 1s, the decimal value is 255.

Visão Geral:

When saving to ASCII format, select the channels to be saved or select All. When channels are selected, only those channels will be saved. When All is selected, all channels are saved.

Visão Geral:

Digital Scope DL708/DL708E: The ON/OFF conditions of bit signals from A1 to A8 and B1 to B8 are expressed as a 16-bit value. The 16-bit value is then converted to a decimal value and the result is saved. The value is written as a ...

Visão Geral:

The dead time is approximately 10µs.

Visão Geral:

During the averaging mode, the operation of the Single Start key differs from the usual operation. Exponential averaging: Operates in the same fashion as when the START/STOP key is pressed. The operation does not stop ...

Visão Geral:

Even though the display record length is shorter than the specified record length, measurement is made over the full record length. There is no function that allows you to move the display position to view the other ...

Visão Geral:

The SINGLE START key is active only when the measurement is stopped and not in progress. Pressing the SINGLE START key or the START/STOP key will stop measurements in progress. This applies to all DL Series models ...

Visão Geral:

The following statement "setup time (15 ns typical) with respect to the edge of the logic trigger" referes to the amount of time in which the combination conditions of the logic input must be met before the detection ...

Visão Geral:

There is zero delay because the external trigger circuit on the DL2700 Digital Oscilloscope is located before the A/D converter. The timing is approximately 10 ns faster than the trigger captured on the measured waveform, since there are ...

Visão Geral:

The data transfer rates for binary data is as follows:                          1 kW 0.035s 10 kW 0.180s 50 kW 0.779s 100 kW 1.536s 250 kW 3.812s 500 kW 7.527s   1 ...

Visão Geral:

The following procedure will allow the hard disk to be read properly. For Windows 95, open the "Settings" tab in the properties dialog box for the connected drive, and check the Removable box. Select "My Computer -> ...

Visão Geral:

The Time Axis Accuracy is the accuracy of the A/D clock. It is not possible to directly measure the A/D clock and the A/D clock is not output from the ScopeCorder DL850/SL1400. To verify the accuracy of the A/D clock, you will ...

Visão Geral:

Question:If the DL9000 Oscilloscopes is in normal trigger mode and no waveform acquisitions have been made i.e. the scope has not triggered, a query to the instrument using the :Waveform:Record or :History:Record? Minimum it returns ...

Visão Geral:

The free software DL-GATE can be used for the following products only:DL1700E seriesDL7400 series (firmware version after 1.32)DL1600 series (firmware version after 1.13)DL750 series (firmware version after ...