How Much Lag Time is There When Inputting or Outting a Signal from the EXT Terminal of the WE7000 Measurement Station?

The lag time from the external terminal to the bus trigger / time base is as follows:

When inputting from TRIG IN:
TRIG IN -> bus trigger signal : 40 ns (typical value)
TRIG IN -> time base signal : 25 ns (typical value)

When inputting from EXT. I/O:
EXT. I/O -> bus trigger signal : 25 ns (typical value)
EXT. I/O ->time base signal : 10ns (typical value)

When outputting to EXT. I/O:
EXT. I/O <- bus trigger signal : 25ns (typical value)
EXT. I/O <- time base signal : 25ns (typical value)

The lag time between the bus trigger / time base and each measurement module differs depending on the module. See the following example:

Bus trigger signal -> WE7111 Digital oscilloscope module : 30 ns
+ sampling interval (typical value*)
* Typical value represents a typical or average value. It is not strictly guaranteed.
 

Related Products & Solutions

WE7000 PC-Based Measurement Instruments

One system, multiple instruments:  WE7000 satisfies demands for fast, reliable and precise data acquisition which uses a standard laptop or PC as its user interface. Input modules plug into an expandable measuring station. Via the optical interface, it is possible to galvanically isolate the laptop/PC.

Data Acquisition (DAQ)

Yokogawa data acquisition systems give you the most flexibility and power to measure, display, store, and even actuate any number of physical or electrical phenomena.

High Speed Data Acquisition

  • Yokogawa high-speed data acquisition systems
  • Industry-leading isolation, bit resolution, sampling rate, and memory depth
  • Independent channel hardware and easy to use software

Precision Making

Top