Why Does the DLM2000 Mixed Signal Oscilloscope Save Only One Acquisition Waveform Even Though History Mode is set to ALL?

The DLM2000 Mixed Signal Oscilloscope can save all history acquisition waveforms into one file, if the data type is set to Binary (file extension *.WDF) and History Mode is set to All. If the data type is set to ASCII, only the latest acquisition waveform will be saved, regardless if History Mode is set to All or One. When saving Binary history files, please make sure that the Sampling Mode, in the ACQUIRE menu, is set to RealTime.

Related Products & Solutions

DLM2000 Mixed Signal Oscilloscopes

200, 350, and 500MHz mixed-signal oscilloscopes for every engineer. Best-in-class performance in usability, acquisition, analysis, and display—all at a price you can digest. Options include serial bus, vehicle bus, and power supply analysis functions.

Mixed Signal Oscilloscopes

  • Yokogawa mixed signal oscilloscopes
  • Simultaneous, time correlated observations
  • Analysis of analog with digital (logic) signals
  • Troubleshooting electrical anomalies, measuring parametric values, monitoring signals

Oscilloscopes

  • Digital oscilloscopes
  • High-speed sampling
  • Range of bandwidths for electronic device design and development
  • Advantages of oscilloscope and multi-channel data recorder

Precision Making

Top