Unable to Re-Connect to DL850 ScopeCorder via GPIB after a WDF File is Loaded Into Acquisition Memory

Unable to connect to DL850 ScopeCorder via GPIB after a WDF file is loaded into acquisition memory.
 
This is a confirmed issue and exists only when using GPIB.
(Ethernet and USB do not exhibit this behaviour.)
 
Solution: Please upgrade your DL850 to version 2.05 or later firmware.
 
 