Can the Width of the Harmonic Analysis Window be Increased Beyond 16?

In order to analyze the harmonics by the PLL synchronization, the maximum possible window width is automatically determined depending on the frequency of the fundamental wave of the input signal (fundamental frequency f).

When the fundamental frequency f of the input signal is 40 =/< f < 70 Hz, the width cannot be set higher than 16. The following are the settings possible, given different fundamental frequencies f:

  • 10 =/< f < 20 Hz: 1 to 4
  • 20 =/< f < 40 Hz: 1 to 8
  • 40 =/< f < 70 Hz: 1 to 16
  • 70 =/< f < 130 Hz: 1 to 32
  • 130 =/< f < 250 Hz: 1 to 64
  • 250 =/< f =/< 440 Hz: 1 to 128
 

Related Products & Solutions

WE7000 PC-Based Measurement Instruments

One system, multiple instruments:  WE7000 satisfies demands for fast, reliable and precise data acquisition which uses a standard laptop or PC as its user interface. Input modules plug into an expandable measuring station. Via the optical interface, it is possible to galvanically isolate the laptop/PC.