What is the Response Time of the 701260 Analog Voltage Module When in RMS Observation Mode?

The measured output response time of the 701260 Analog Voltage Input Module, when observing RMS is:
  • Rising (0 to 90% of 10 div): 100ms
  • Falling (100 to 10% of 10 div): 250ms
An attached JPEG displays the results of the 701260 output response to a step-input with a 60 Hz sine wave. Two zoom windows are displayed, the bottom-left zoom windows is the step-down response and the right window is the step-up response. The response time can be measured only when observing RMS.  

Related Products & Solutions

Data Acquisition (DAQ)

Data acquisition (DAQ) is the measurement, recording, analyzing, and presentation of real world phenomena. It includes electrical measurements such as voltage, current, power as well as measurements through sensors and transducers including temperature, pressure, strain, vibration, and more.

High Speed Data Acquisition

Yokogawa high speed data acquisition systems deliver industry leading isolation, bit resolution, sampling rate, and memory depth, with independent channel hardware and easy to use software.

Oscilloscopes

Capture, analyze, and troubleshoot electrical or real-world physical signals. Oscilloscopes observe the change of electrical signals over time, continuously graphed on a display as voltage or amplitude vs. time.

ScopeCorders

A ScopeCorder is an instrument combining a mixed signal oscilloscope and portable data acquisition recorder into a modular platform designed to capture both high-speed transients and low speed trends. Yokogawa’s ScopeCorder product family provides flexible and high-performance multi-channel test instruments by combining a variety of signal conditioning input modules, on-board calculations, and deep data acquisition storage into a off-the-shelf data acquisition solution.

Top