What is the Response Time of the 701260 Analog Voltage Module When in RMS Observation Mode?

The measured output response time of the 701260 Analog Voltage Input Module, when observing RMS is:
  • Rising (0 to 90% of 10 div): 100ms
  • Falling (100 to 10% of 10 div): 250ms
An attached JPEG displays the results of the 701260 output response to a step-input with a 60 Hz sine wave. Two zoom windows are displayed, the bottom-left zoom windows is the step-down response and the right window is the step-up response. The response time can be measured only when observing RMS.  

Related Products & Solutions

Data Acquisition (DAQ)

Yokogawa data acquisition systems give you the most flexibility and power to measure, display, store, and even actuate any number of physical or electrical phenomena.

High Speed Data Acquisition

Yokogawa high speed data acquisition systems deliver industry leading isolation, bit resolution, sampling rate, and memory depth, with independent channel hardware and easy to use software.

Oscilloscopes

World-class digital oscilloscopes from Yokogawa: The digital oscilloscopes have high-speed sampling and a wide range of bandwidths that can be utilized for design and development of electronic devices. The ScopeCorders have the advantages of both a digital oscilloscope and a multi-channel data recorder.

ScopeCorders

A ScopeCorder is an instrument combining a mixed signal oscilloscope and portable data acquisition recorder into a modular platform designed to capture both high-speed transients and low-speed trends. Yokogawa’s ScopeCorder product family provides flexible and high-performance multi-channel test instruments by combining a variety of signal conditioning input modules, onboard calculations, and deep data acquisition storage into an off-the-shelf data acquisition solution.

Top