How to Display Averaged Power Value During Integration on the WT1600 Digital Power Analyzer?

When the WT1600 Digital Power Analyzer is set into Integration mode, the averaged power (watt) values can be calculated and displayed. This is available only by using the User-Defined Function feature found in the MEASURE button menu.
 
The following equation must be entered in any one of the twenty Expression field:
WH(E1)/TI(E1)*3600
 
To display the calculated result, simply select and display the User Defined Function labeled F1 to F2 in the ITEM button menu.
 
The averaged power equation is pre-set on the WT500 and WT3000.

Produtos e Soluções Relacionadas

Precision Power Analyzer WT3000

With 0.02% accuracy and 1MHz bandwidth, the WT3000 delivers where the highest precision measurements are required. It is the industry standard for R&D work on inverters, motor drives, lighting systems and electronic ballasts, UPS systems, aircraft power, transformer testing, and other power conversion devices.

WT1600 Digital Power Meter

The High-end WT1600 is designed for a wide range of applications, from energy-saving to large load applications. The WT1600 has a six element input capability to provide maximum flexibility, and Basic Power Accuracy of 0.1% with 1 MHz Bandwidth.

WT500 - Mid Range

  • Power and Harmonic (THD) Measurements with Independent Range Controls
  • Efficiency measurements of AC/DC or DC/AC Inverters & Motors

 

Analisadores de Potência e Medidores de Potência

A Yokogawa, a maior fabricante mundial de analisadores e medidores de energia e energia, fornece uma ampla gama de analisadores de energia digital para atender a todos os requisitos.

Precision Making

Topo