How to use SNTP on DL850 ScopeCorder
The sequence below is very important for SNTP to work properly.
1. Set DATE/TIME and GMT Time Difference to your location FIRST. EDT is -4 hours, EDT is -5 hours - as examples.
2. Set a SNTP Server for GMT in SNTP settings (64.90.182.55 in New York City for example - this is GMT, not EDT or EST);
3. Set up TCP IP as usual.
4. Under SNTP function, press ADJUST .. this must be performed as the last step.