What is the Step Response of the 701280 Frequency Input Module for a Sine Wave Filtered with an IIR LP Filter?

For the 701280 Frequency Input Module with a 4th order IIR 6Hz lowpass filter, a 60Hz sine wave (FM modulated with a square wave) was measured with deviations of ± 3Hz. Please download and view the attached image for a screenshot of multiple measurements.

The top waveform is the output of the 701280 module driven with a 60Hz unit step deviated at ± 3Hz. The 701280 module is measuring in Frequency mode.

 
The middle waveform is the top waveform after an IIR 6Hz LP filter has been applied.
 
The bottom waveform is the 60Hz input sine wave.

Produtos e Soluções Relacionadas

Aquisição de Dados (DAQ)

Os sistemas de aquisição de dados da Yokogawa proporcionam a você mais flexibilidade e poder para medir, exibir, armazenar e até mesmo atuar qualquer número de fenômenos físicos ou elétricos.

High Speed Data Acquisition

  • Yokogawa high-speed data acquisition systems
  • Industry-leading isolation, bit resolution, sampling rate, and memory depth
  • Independent channel hardware and easy to use software

Osciloscópios

Osciloscópios digitais de classe mundial da Yokogawa: Os osciloscópios digitais têm amostragem de alta velocidade e uma ampla gama de larguras de banda que podem ser utilizadas para projeto e desenvolvimento de dispositivos eletrônicos. As ScopeCorders têm as vantagens de um osciloscópio digital e de um gravador de dados multicanal.

OSCILÓGRAFOS

  • Flexible, high-performance Yokogawa ScopeCorder
  • Modular platform combines mixed signal oscilloscope and portable data acquisition recorder 
  • Captures high-speed transients and low-speed trends 

 

Precision Making

Topo