Why WT210/230 and WT3000 can not calculate Urms, Unm, and Udc simultaneously?

There are restrictions of the DSP hardware.
 
WT230 and WT210 are low-cost power meters.
So, a low-cost, modest performance DSP utilized.
This DSP has slow calculation.
 
WT3000 is performing complicated calculation in order to maintain high accuracy.

So WT3000's DSP needs a long time to calculate the data.
 
If a lot of data is calculated simultaneously, it will not be in time for the update time of measurement.
There was only one way to calculate only one mode Urms, Umn or Udc.
So, WT210, WT230 and WT3000 can not calculate Urms, Unm, and Udc simultaneously.
PZ4000, WT1600, WT1800 and WT500 have a high speed DSP.
So they can calculate Urms, Unm, and Udc simultaneously.

Produtos e Soluções Relacionadas

Precision Power Analyzer WT3000

With 0.02% accuracy and 1MHz bandwidth, the WT3000 delivers where the highest precision measurements are required. It is the industry standard for R&D work on inverters, motor drives, lighting systems and electronic ballasts, UPS systems, aircraft power, transformer testing, and other power conversion devices.

Analisadores de Potência e Medidores de Potência

A Yokogawa, a maior fabricante mundial de analisadores e medidores de energia e energia, fornece uma ampla gama de analisadores de energia digital para atender a todos os requisitos.

Precision Making

Topo