In recent years there have been increasing deployments of the fiber to the home (FTTH) service in access network areas, along with advances in transmission technologies such as B-PON, G-PON, and GE-PON. In addition, there have been important advances in CWDM transmission technology, which is expected to be used in metropolitan areas. As networks have evolved, there has been an increasing need for high price-performance measuring instruments capable of high-throughput evaluation of the optical characteristics of optical devices and components used in these facilities. The AQ2200-136 TLS module is an optional wideband variable wavelight source designed for the A2200 multi-application test system. Although it is lightweight with a small, 2-slot module, it is capable of tuning the range of 1440 nm to 1640 nm.
The AQ2200-136 TLS module is an optional wideband variable wavelength light source designed for the AQ2200 multi-application test system.
Swept measurement synchronized with optical spectrum analyzer (standard function)
Swept measurement system (standard function)
Parameter | Product specification | Remarks | |
---|---|---|---|
Wavelength varying range 1440 to 1640 nm | |||
Wavelength resolution | 1 pm | ||
Absolute wavelength accuracy | ≤±100 pm | 1 | |
Relative wavelength accuracy | ≤±50 pm | 1 | |
Wavelength stability | ±5 pm/h | 2, 4, 5 | |
Spectral line width | NARROW | ≤1 MHz | 2, 4, 8 |
WIDE | ≥50 MHz | 1, 2, 4, 8 | |
Optical output | Maximum output wavelength | ≥±7 dBm | 1, 10 |
1520 to 1610 nm | ≥±5 dBm | ||
Total wavelength range | ≥-8 dBm | ||
Optical output stability | 5 minutes | ≤±0.01 dB | 1, 3, 5, 6, 10 |
1 hour | ≤±0.05 dB | ||
24 hours | ≤±0.1 dB | ||
Optical output repeatability | ≤±0.04 dB | 1, 3, 5, 6, 10 | |
Monitor output | Maximum optical output | ≤±5 dBm | 9 |
Minimum optical output | ≥-25 dBm | ||
Accept optical fiber | SM (ITU-T G.652) | ||
Accepted optical connector | FC/Angled PC | 7 |
Laser safety information
Unless otherwise noted, the specifications represent values obtained after calibration performed at least one hour OPTON, under reference operatingconditions*with 2 m SMF output point.
*Reference operating conditions: Ambient temperature of 23±2ºC, temperature change rates of 1ºC per 10 minutes and 3ºC per hour, and humidity of 50±5%
1At maximum optical output
2Wavelength: 1590 nm
3Wide line width
4Narrow line width
5Constant temperature
6Wavelength: 1460 to 1620 nm
7SEIKOH GIKEN Angled/PC or conforming product (step type)
8At minimum optical output
9Using MONITOR connector
10Using OUTPUT connector