Delta Computation Function

It is possible to obtain the differential voltage, line voltage, phase voltage, etc. by obtaining the sums and differences of instantaneous measurement values of voltage and current in each element.

Delta Image
  • Differential voltage/current: Differential voltage and current between two elements are computed in the three-phase three-wire system.

  • Line voltage/phase current: Line voltage and phase current that are not measured are computed in the three-phase three-wire system (Figure 1).
  • Star-delta conversion: Line voltage is computed from the phase voltage using the three-phase four-wire system data.
  • Delta-star conversion: Phase voltage is computed from the line voltage in the three-phase three-wire system (3V3A system)
    (Figure 2).
DeltaBu1
Figure 1. Line voltage/phase current
  DeltaBu2
Figure 2. Delta-star conversion

Related Products & Solutions

WT1800E High Performance Power Analyzer

The WT1800E is a flexible and reliable power analyzer that guarantees power accuracy of 0.05% of reading plus 0.05% of range. It is capable of harmonics analysis up to the 500th order of a 50/60 Hz fundamental frequency. With up to 6 input channels, a wide range of display and analysis features, and PC connectivity, the WT1800E can handle broad variety of power efficiency and harmonic analysis requirements.

 
Top