What is the Correct AC Coupling Test for the 720210 Analog Voltage Input Module?

The correct AC coupling test for the 720210 Analog Voltage module must be performed at 20 Hz.
 
The service manual, SMDL850-01EN_010 1st Ed, suggests that you perform the AC COUPLING test at 15 Hz. This is a typo and incorrect, the test must be performed at 20 Hz.

Additional Information
  • Confirmation: The QIS720210-E_010 has the correct procedure outline for the AC COUPLING TEST at 20Hz.
  • Confirmation: Yokogawa Corp of America Cal Lab using Intercal METCAL script performs this test at 20 Hz.

Related Products & Solutions

SL1000 High Channel Count ScopeCorder

A headless ScopeCorder system capable of connecting multiple chassis into one PC-based system.

 

Data Acquisition (DAQ)

Yokogawa data acquisition systems give you the most flexibility and power to measure, display, store, and even actuate any number of physical or electrical phenomena.

High Speed Data Acquisition

Yokogawa high speed data acquisition systems deliver industry leading isolation, bit resolution, sampling rate, and memory depth, with independent channel hardware and easy to use software.

Oscilloscopes

World-class digital oscilloscopes from Yokogawa: The digital oscilloscopes have high-speed sampling and a wide range of bandwidths that can be utilized for design and development of electronic devices. The ScopeCorders have the advantages of both a digital oscilloscope and a multi-channel data recorder.

ScopeCorders

A ScopeCorder is an instrument combining a mixed signal oscilloscope and portable data acquisition recorder into a modular platform designed to capture both high-speed transients and low-speed trends. Yokogawa’s ScopeCorder product family provides flexible and high-performance multi-channel test instruments by combining a variety of signal conditioning input modules, onboard calculations, and deep data acquisition storage into an off-the-shelf data acquisition solution.

Top