Applicable model*: AQ6380/AQ6374/AQ6375/AQ6376/AQ6377
When a light source with a wide wavelength band is measured with an optical spectrum analyzer (OSA), noisy waveforms may appear in certain wavelength bands and interfere with measurement. This is due to light absorption of water vapor (H2O) or carbon dioxide (CO2) contained in the air inside the OSA optics (monochromator). Water vapor has strong optical absorption near 1400 nm, 1900 nm and 2700 nm, and carbon dioxide has near 2000 nm and 4200 nm.
The applicable models are equipped with a purging mechanism to address this issue. By continuously supplying and exhausting purge gas such as dry air and nitrogen from the supply and exhaust port on the rear panel to the inside of the monochromator, the water vapor and carbon dioxide inside the OSA are removed, and the correct optical spectrum can be measured.
Purging gas ports
(IN/OUT)
Major requirements for purging nSupply tubing
1/4” outer diameter quick connectors, clean flexible nylon tubing nMaximum rating
Pressure: 1.5 psig (0.01 MPaG) nPurge gas
Ultra-high purity liquid nitrogen which must be 99.9999% pure grade with the heat exchanger is recommended. |
* All versions unless otherwise specified.
For AQ6375, AQ6375B or later.
No purging
Nitrogen purging (1 hr.)
Replacement efficiency of purging
High Performance LONG WAVELENGTH
The AQ6375B is a bench-top optical spectrum analyzer covering the long wavelengths, 1200 to 2400 nm, with the added benefits of gas purging input ports / output ports, a built-in cut filter for high order diffracted light, and a novel double speed mode which increases the sweep speed up to 2 times compared to the standard sweep mode.
High Performance LONG WAVELENGTH
The AQ6375E covers not only telecommunication wavelengths, but also the SWIR region which is often used for environmental sensing and medical applications.
・Lineup of 3 models [Standard, Extended and Limited]
・Covers wavelengths
1200 to 2400 nm [Standard, and Limited]
1000 to 2500 nm [Extended ]
MWIR WAVELENGTH with internal gas purge and cut filter
The AQ6376 is the latest version of our bench-top optical spectrum analyzer extending the wavelength coverage well beyond the NIR range of our previous models into the MWIR region from 1500 to 3400 nm.
Popular applications include the detection of gases such as carbon oxides (COx), nitrogen oxides (NOx), and hydrocarbon gas (CxHy) for environmental studies.
MWIR WAVELENGTH with internal gas purge and cut filter
The AQ6376E is the latest version of our bench-top optical spectrum analyzer extending the wavelength coverage well beyond the NIR range of our previous models into the MWIR region from 1500 to 3400 nm.
Popular applications include the detection of gases such as carbon oxides (COx), nitrogen oxides (NOx), and hydrocarbon gas (CxHy) for environmental studies.
AQ6380 OSA: 5 pm high wavelength resolution, ±5 pm accuracy, 65 dB wide close-in dynamic range, 80 dB high stray light suppression