When a light source with a wide wavelength band is measured with an optical spectrum analyzer (OSA), noisy waveforms may appear in certain wavelength bands and interfere with measurement. This is due to light absorption of water vapor (H2O) or carbon dioxide (CO2) contained in the air inside the OSA optics (monochromator). Water vapor has strong optical absorption near 1400 nm, 1900 nm and 2700 nm, and carbon dioxide has near 2000 nm and 4200 nm.
The applicable models are equipped with a purging mechanism to address this issue. By continuously supplying and exhausting purge gas such as dry air and nitrogen from the supply and exhaust port on the rear panel to the inside of the monochromator, the water vapor and carbon dioxide inside the OSA are removed, and the correct optical spectrum can be measured.
Purging gas ports
(IN/OUT)
Major requirements for purging nSupply tubing
1/4” outer diameter quick connectors, clean flexible nylon tubing nMaximum rating
Pressure: 1.5 psig (0.01 MPaG) nPurge gas
Ultra-high purity liquid nitrogen which must be 99.9999% pure grade with the heat exchanger is recommended. |
Applicable model*: AQ6374/AQ6375/AQ6376/AQ6377
* For AQ6375, AQ6375B or later.
Der optische Breitband-Spektrumanalysator AQ6374 deckt den Wellenlängenbereich von 350 bis 1750 nm und damit sowohl den Bereich des sichtbaren Lichts (380 bis 780 nm) als auch die im Telekommunikationsbereich genutzten Wellenlängen ab.
Hohe Leistung bei großen Wellenlängen
Der optische Benchtop-Spektrumanalysator AQ6375B deckt den Wellenlängenbereich von 1200 bis 2400 nm ab. Gegenüber dem Vorgängermodell bietet er eine kontinuierliche Gasspülung der Ein- und Ausgänge, einen integrierten steilflankigen Filter für gebeugtes Licht höherer Ordnung und einen neuartigen Messmodus, der die Sweep-Geschwindigkeit gegenüber dem Standard-Sweep-Modus verdoppelt.
High Performance LONG WAVELENGTH
The AQ6375E covers not only telecommunication wavelengths, but also the SWIR region which is often used for environmental sensing and medical applications.
・Lineup of 3 models [Standard, Extended and Limited]
・Covers wavelengths
1200 to 2400 nm [Standard, and Limited]
1000 to 2500 nm [Extended ]
MWIR-Wellenlängenbereich mit interner Gasspülung und Tiefpass-Filter
Der AQ6376 ist die neuste Ergänzung unserer optischen Benchtop-Spektrumanalysatoren und erweitert die Wellenlängen-Abdeckung weit über den NIR-Bereich der anderen Modelle hinaus in den MWIR-Bereich von 1.500 bis 3.400 nm.
MWIR WAVELENGTH with internal gas purge and cut filter
The AQ6376E is the latest version of our bench-top optical spectrum analyzer extending the wavelength coverage well beyond the NIR range of our previous models into the MWIR region from 1500 to 3400 nm.
Popular applications include the detection of gases such as carbon oxides (COx), nitrogen oxides (NOx), and hydrocarbon gas (CxHy) for environmental studies.
Yokogawa bietet Lösungen für Messungen sowohl an konventionellen, als auch an modernsten optischen Komponenten und Systemen für die Nachrichtentechnik und die allgemeine Optik an. Derartige Messungen werden besonders bei optischen Fasern mit hoher Kapazität und neuen Bauteiltechnologien immer wichtiger. Bei Yokogawa wird dabei ein Wellenlängenbereich vom sichtbaren bis in den infraroten Bereich um 2,5µm abgedeckt.
Yokogawa gehört mit der Produktlinie der hochwertigen und innovativen optischen Spektrumanalysatoren (OSA) zu den weltweiten Marktführern bei der optischen Wellenlängen-Messtechnik. Auf der Basis unserer mehr als dreißigjährigen Erfahrungen in diesem Bereich ergänzen wir unsere OSAs nun mit optischen Wellenlängen-Messgeräten als komplementäre Lösung. Die optischen Wellenlängen-Messgeräte von Yokogawa erlauben schnelle, genaue und dennoch kostengünstige optische Wellenlängen-Messungen entsprechend den derzeitigen und künftigen Anforderungen und im Hinblick auf den weltweit schnell zunehmenden Datenverkehr in den Netzen.