AQ23295A Optical Sensor Head (Si)

Large-Area Si Sensor Head for Short-Wavelength Optical Power Measurements 

The AQ23295A is an optical sensor head for measuring optical power over the short-wavelength range from 400 to 1100 nm. Its large-area Si detector supports stable measurements of light from single-mode fibers, multimode fibers, and free-space optical sources.

The handheld sensor head can be positioned close to the device under test, making it suitable for optical device evaluation, multi-fiber connector testing, and measurements in setups where direct access to the optical signal is required. The AQ23295A is used with the AQ23202A Interface Module in an AQ2300 Series Test Platform.

Key Specifications 

  • Detector type: Si □5.8 × 5.8 mm 
  • Power range: -90 to +10 dBm (CW) 
  • Wavelength range: 400 to 1100 nm 
  • Uncertainty under reference conditions: ±2.5% 
  • Averaging time (minimum): 20 µs 
  • Applicable fiber: SMF, MMF 
  • Connector: FC, SC, LC, MU (MPO is also available as an optional accessory.)
  • Cable length: 1 m, 4 m 

Specifications

AQ23295A Optical Sensor Head (Si)

Items Specifications
ModelAQ23295A
Number of channels1 channel
Detector typeSi, □5.8 mm diameter
Wavelength range400 to 1100 nm
Power range (CW)−90 to +10 dBm
Minimum display resolution1/10000
Applicable fiber≤62.5/125 µm (GI), NA ≤0.275
Uncertainty under reference conditions±2.5%*2, *4, *5
Total uncertainty±5% ±5 pW*2, *7, *8
Polarization dependence—
Linearity±0.04 dB ±5 pW*2, *11, *12, *14
Noise level5 pW or less*16
Averaging time (minimum)20 µs
Optical connectorAQ9335C connector adapter*17
Calibration cycle1 year
Dimensions (excluding protrusions)106.5 (H) × 31 (W) × 321.5 (D) mm
WeightApprox. 0.3 kg
Note:
*All values in the specifications assume a warm-up period of one hour.
*The environmental conditions are subject to the specification of frame controller.
*1: At 1310 nm
*2: When the AQ9335C Connector Adapter is used.
*3: Wavelength: 1310/1550±20 nm, optical fiber: SM (ITU-T G.652.D) NA = 0.1, polarization dependence is not included
*4: Wavelength: 850±15 nm, optical fiber: SM (850 nm SMF)
*5: Power level: 100 μW (‒10 dBm), CW light, light source spectrum width: 10 nm or less, ambient temperature: 23±2°C, optical connector: FC/PC, averaging: 1 second, wavelength setting error: 0.5 nm or less, changes to the measuring instrument due to the passage of time are not included in these conditions. More than 1 year but less than 2 years since the last calibration: add 0.4%, over 2 years: add 0.8%.
*6: Wavelength: 1000 to 1630 nm, polarization dependence is not included
*7: Wavelength: 500 to 900 nm (900 to 1000 nm: add ±2%)
*8: Power level: 1 nW to 10 mW (‒60 dBm to +10 dBm), CW light, light source spectrum width:10 nm or less, ambient temperature: 23±5°C, optical fiber: GI50 NA0.2 (add ±1% if GI62.5 and NA0.275), optical connector: FC/PC, averaging: 1 second, auto range, wavelength setting error: 0.5 nm or less, changes to the measuring instrument due to the passage of time are not included in these conditions. More than 1 year but less than 2 years since the last calibration: add 0.4%, over 2 years: add 0.8%.
*9: Wavelength: 1550 nm, ambient temperature: 23±2°C, optical fiber: SM (ITU-T G.652.D) NA = 0.1, optical connector: FC/PC
*10: Wavelength: 1000 to 1630 nm, optical fiber: SM (ITU-T G.652) NA = 0.1
*11: Wavelength: 500 to 900 nm, beam condition: SM (850 nm SMF), GI50 NA = 0.2 or GI62.5 NA = 0.275
*12: Power level: 1 nW to 10 mW (‒60 dBm to +10 dBm), CW light, ambient temperature: 23±2°C (constant temperature), auto range, averaging: 1 second, optical connector: FC/PC
*13: For free-space light measurement, the power density of the light-detecting surface is 30 mW/mm2 or less.
*14: For free-space light measurement, the power density of the light-detecting surface is 5 mW/mm2 or less.
*15: Wavelength: 1000 to 1630 nm, ambient temperature: 23±2°C (constant temperature), averaging: 1 second, within 5 minutes after zero set execution
*16: Wavelength: 500 to 900 nm, ambient temperature: 23±2°C (constant temperature), averaging: 1 second, within 5 minutes after zero set execution
*17: Select from FC, SC, LC, or MU (It is also possible to measure free-space light by removing the connector adapter.)