What is the True Max DC Input Voltage for the DLM2000 Mixed Signal Oscilloscope?

The actual input limit for the DLM2000 Mixed Signal Oscilloscope is 212V (DC+ACpeak) for not only a 0.01 or 0.001 Hz AC signal but also a DC signal. The reason why 150 Vrms is indicated on the DLM2000 specification sheet is that the safety standard requires the max input voltage to be defined by units of rms.

Produtos e Soluções Relacionadas

DLM2000 Mixed Signal Oscilloscopes

200, 350, and 500MHz mixed-signal oscilloscopes for every engineer. Best-in-class performance in usability, acquisition, analysis, and display—all at a price you can digest. Options include serial bus, vehicle bus, and power supply analysis functions.

Mixed Signal Oscilloscopes

  • Yokogawa mixed signal oscilloscopes
  • Simultaneous, time correlated observations
  • Analysis of analog with digital (logic) signals
  • Troubleshooting electrical anomalies, measuring parametric values, monitoring signals

Osciloscópios

Osciloscópios digitais de classe mundial da Yokogawa: Os osciloscópios digitais têm amostragem de alta velocidade e uma ampla gama de larguras de banda que podem ser utilizadas para projeto e desenvolvimento de dispositivos eletrônicos. As ScopeCorders têm as vantagens de um osciloscópio digital e de um gravador de dados multicanal.