When Adjusting a WE Module With the Adjustment Software, Which Slot do you use to Install the Module?

When using the Adjustment Software to adjust a WE7000 Pc Based Measurement Instruments input module, the module should be installed in slot 2 of the main station. The slot labels on the WE Main Station begin with "0". Therefore, install the module in slot labelled as "2".
 
Also note that you can adjust only one module at a time. Therefore, there should no other modules installed in the main station, except the module that is to be adjusted, installed in slot "2".
 
Otherwise the adjustment software will not work.

Produtos e Soluções Relacionadas

WE7000 PC-Based Measurement Instruments

One system, multiple instruments:  WE7000 satisfies demands for fast, reliable and precise data acquisition which uses a standard laptop or PC as its user interface. Input modules plug into an expandable measuring station. Via the optical interface, it is possible to galvanically isolate the laptop/PC.

Aquisição de Dados (DAQ)

Os sistemas de aquisição de dados da Yokogawa proporcionam a você mais flexibilidade e poder para medir, exibir, armazenar e até mesmo atuar qualquer número de fenômenos físicos ou elétricos.