Why Does the Self Test Fail on the WE7111 Digital Oscilloscope Module?

You must use a probe with a attenuator of "10:1", otherwise the self test will not run correctly. Also, before executing the self-test, please allow for a 30 minutes warm-up time. If the instrument does not warm up, the self test may not be accurate.

Related Products & Solutions

Data Acquisition (DAQ)

Scalable DAQ systems with industry-leading isolation, noise immunity, built-in conditioning, and real-time analysis, ensuring accurate, reliable measurements and faster decisions.

High Speed Data Acquisition

  • PC-based, streaming, local, or remote operation
  • 20+ modules, isolated and versatile inputs
  • Up to 200 MS/s or 640 ch
  • Used in aerospace, automotive, energy, and manufacturing industries
  •  

WE7000 PC-Based Measurement Instruments

One system, multiple instruments:  WE7000 satisfies demands for fast, reliable and precise data acquisition which uses a standard laptop or PC as its user interface. Input modules plug into an expandable measuring station. Via the optical interface, it is possible to galvanically isolate the laptop/PC.