If during the DL850/DL850 firmware upgrade process, the display screen turns black, blue, or something in between and the installation failed, please complete the following steps. You will need a USB printer cable to complete the recovery process.
  1. Download the file
    http://support.us.yokogawa.com/downloads/TMI/HFI/DL850/DL850_Recovery_to_110_via_USB_Cable.zip
  2. Contact Yokogawa Technical Support with the unit's serial number and for zipped file password
  3. Unzip all files to a folder destination
  4. Connect the DL850 to PC with a USB printer cable
  5. Install the USB driver onto your PC
  6. Power DL850 OFF
  7. While holding the F2 soft key, power DL850 ON
  8. Wait for at least 15 seconds
  9. Double click the Update850.BAT file
  10. Follow on-screen prompts
  11. When binary upload to DL850 is complete, cycle the power
  12. Power DL850 ON
  13. Allow firmware installation to complete

Related Products & Solutions

Data Acquisition (DAQ)

Yokogawa data acquisition systems give you the most flexibility and power to measure, display, store, and even actuate any number of physical or electrical phenomena.

High Speed Data Acquisition

Yokogawa high speed data acquisition systems deliver industry leading isolation, bit resolution, sampling rate, and memory depth, with independent channel hardware and easy to use software.

Oscilloscopes

World-class digital oscilloscopes from Yokogawa: The digital oscilloscopes have high-speed sampling and a wide range of bandwidths that can be utilized for design and development of electronic devices. The ScopeCorders have the advantages of both a digital oscilloscope and a multi-channel data recorder.

ScopeCorders

A ScopeCorder is an instrument combining a mixed signal oscilloscope and portable data acquisition recorder into a modular platform designed to capture both high-speed transients and low-speed trends. Yokogawa’s ScopeCorder product family provides flexible and high-performance multi-channel test instruments by combining a variety of signal conditioning input modules, onboard calculations, and deep data acquisition storage into an off-the-shelf data acquisition solution.

Top