DL850 ScopeCorder - How to maximize repeatability of Phase Measurements

DL850 ScopeCorder - How to maximize repeatability of Phase Measurements
 
  1. Use a direct-input if possible; this will remove the delay due to passive-probes 700929 being different between a pair of input channels (in phase response);
  2. Be sure INPUT COUPLING is set properly; it is easy to overlook; so, if you do use an input probe, be sure to set to 10:1, as an example.
  3. Set your V/DIV or SPAN up so as to maximize VERTICAL RESOLUTION - us as near to 20 vertical divisions as possible; this will decrease the influence of noise on the zero-crossing measurement of each waveform.
  4. Optional: using a BW-limiting filter at the front-end would further improve stability; it may actually (in theory) introduce some detectable phase-shift, so be sure to use the same filter on both channels if filtering is desirable.

Related Products & Solutions

Data Acquisition (DAQ)

Simplify complex testing with DAQ systems that flex to your needs. Capture and analyze any physical or electrical signal with ease, transforming signals into actionable insights that speed innovation and improve reliability.

High Speed Data Acquisition

  • PC-based, streaming, local or remote operation
  • 20+ modules, isolated and versatile inputs
  • Up to 200 MS/s or 640 ch

Isolated Oscilloscopes | ScopeCorders

  • An integrated measurement system for every electromechanical application
  • Modular platform combines oscilloscope and DAQ functionality 
  • Capture high-speed transients and low-speed trends 

Oscilloscopes

Accelerate design and troubleshooting with oscilloscopes that combine high-speed sampling, long record lengths, and mixed-signal analysis. From quick debug to advanced R&D, Yokogawa oscilloscopes help you see deeper, work faster, and design smarter.

Precision Making

Top