  1. Use a direct-input if possible; this will remove the delay due to passive-probes 700929 being different between a pair of input channels (in phase response);
  2. Be sure INPUT COUPLING is set properly; it is easy to overlook; so, if you do use an input probe, be sure to set to 10:1, as an example.
  3. Set your V/DIV or SPAN up so as to maximize VERTICAL RESOLUTION - us as near to 20 vertical divisions as possible; this will decrease the influence of noise on the zero-crossing measurement of each waveform.
  4. Optional: using a BW-limiting filter at the front-end would further improve stability; it may actually (in theory) introduce some detectable phase-shift, so be sure to use the same filter on both channels if filtering is desirable.

Produtos e Soluções Relacionadas

Aquisição de Dados (DAQ)

Os sistemas de aquisição de dados da Yokogawa proporcionam a você mais flexibilidade e poder para medir, exibir, armazenar e até mesmo atuar qualquer número de fenômenos físicos ou elétricos.

High Speed Data Acquisition

  • Yokogawa high-speed data acquisition systems
  • Industry-leading isolation, bit resolution, sampling rate, and memory depth
  • Independent channel hardware and easy to use software

Osciloscópios

Osciloscópios digitais de classe mundial da Yokogawa: Os osciloscópios digitais têm amostragem de alta velocidade e uma ampla gama de larguras de banda que podem ser utilizadas para projeto e desenvolvimento de dispositivos eletrônicos. As ScopeCorders têm as vantagens de um osciloscópio digital e de um gravador de dados multicanal.

OSCILÓGRAFOS

  • Flexible, high-performance Yokogawa ScopeCorder
  • Modular platform combines mixed signal oscilloscope and portable data acquisition recorder 
  • Captures high-speed transients and low-speed trends 

 

Precision Making

Topo