What is the Accuracy Specification for the 701280 Frequency Module When Operated Outside the Standard Operating Conditions Range?

The 701280 frequency input module measures only time domain values, it does not measure any voltage or any amplitude domain value. Since the temperature coefficients are only applied to the amplitude domain value, the temperature coefficients are not relevant to the 701280 module. The accuracies of 70128 are available in a operating temperature range of 5 to 40°C.

Related Products & Solutions

Data Acquisition (DAQ)

Yokogawa data acquisition systems give you the most flexibility and power to measure, display, store, and even actuate any number of physical or electrical phenomena.

High Speed Data Acquisition

Yokogawa high speed data acquisition systems deliver industry leading isolation, bit resolution, sampling rate, and memory depth, with independent channel hardware and easy to use software.

Oscilloscopes

World-class digital oscilloscopes from Yokogawa: The digital oscilloscopes have high-speed sampling and a wide range of bandwidths that can be utilized for design and development of electronic devices. The ScopeCorders have the advantages of both a digital oscilloscope and a multi-channel data recorder.

ScopeCorders

A ScopeCorder is an instrument combining a mixed signal oscilloscope and portable data acquisition recorder into a modular platform designed to capture both high-speed transients and low-speed trends. Yokogawa’s ScopeCorder product family provides flexible and high-performance multi-channel test instruments by combining a variety of signal conditioning input modules, onboard calculations, and deep data acquisition storage into an off-the-shelf data acquisition solution.