What is the Accuracy Specification for the 701280 Frequency Module When Operated Outside the Standard Operating Conditions Range?

The 701280 frequency input module measures only time domain values, it does not measure any voltage or any amplitude domain value. Since the temperature coefficients are only applied to the amplitude domain value, the temperature coefficients are not relevant to the 701280 module. The accuracies of 70128 are available in a operating temperature range of 5 to 40°C.
 