The V Scale SPAN mode on the DL850 ScopeCorder can be used to adjust the upper and lower display limits of the zoom window. To optimize the V Scale SPAN mode to make the best resolution measurements, follow the procedures listed below.
The ScopeCorder modules all have 20 divisions, and not the usual 10 division with standard oscilloscopes. There are actually 25 vertical divisions but we recommend you use only 20 divisions to stay within the specified accuracy for that module.
Note: Make sure to install or update the firmware to version 1.35 or later. Several improvements have been made which includes changes to the SPAN adjustment under any vertical menu/channel.
Please download the attached PowerPoint presentation for more information and detailed measurements regarding the ScopeCorder vertical resolution.
Os sistemas de aquisição de dados da Yokogawa proporcionam a você mais flexibilidade e poder para medir, exibir, armazenar e até mesmo atuar qualquer número de fenômenos físicos ou elétricos.
Osciloscópios digitais de classe mundial da Yokogawa: Os osciloscópios digitais têm amostragem de alta velocidade e uma ampla gama de larguras de banda que podem ser utilizadas para projeto e desenvolvimento de dispositivos eletrônicos. As ScopeCorders têm as vantagens de um osciloscópio digital e de um gravador de dados multicanal.