How do I Obtain the Highest Resolution Measurements Using the V Scale Span Mode on the DL850 ScopeCorder?

The V Scale SPAN mode on the DL850 ScopeCorder can be used to adjust the upper and lower display limits of the zoom window. To optimize the V Scale SPAN mode to make the best resolution measurements, follow the procedures listed below.

  1. Press one of the CH1 to CH16 keys to select the source waveform
  2. Press the Next 1/2 soft button
  3. Press the V Scale soft button and select DIV
  4. Using the SCALE rotary knob, adjust the vertical scale to the most sensitive setting without clipping the waveform
  5. Press the V Scale soft button and select SPAN
  6. Use the Arrow Keys or Jog Shuttle and the SET key to set the Upper and Lower Limit

The ScopeCorder modules all have 20 divisions, and not the usual 10 division with standard oscilloscopes. There are actually 25 vertical divisions but we recommend you use only 20 divisions to stay within the specified accuracy for that module.

Note: Make sure to install or update the firmware to version 1.35 or later. Several improvements have been made which includes changes to the SPAN adjustment under any vertical menu/channel.

Please download the attached PowerPoint presentation for more information and detailed measurements regarding the ScopeCorder vertical resolution.

相关产品和解决方案

数据采集设备

Yokogawa数据采集系统具备出众的灵活性和性能,能测量、显示、存储甚至促动各种物理和电气现象。

波形测量仪器

Yokogawa在业界首推示波记录仪的概念,将数字示波器和多通道数据记录仪的优点集于一身。横河示波记录仪作为世界一流的示波记录仪具有高速采样能力和大测量带宽的优点,可广泛应用于电子工程的设计和开发过程。

横河示波器既有面向低成本市场的4CH MSO示波器DLM3000系列,也有业界唯一的8CH MSO示波器DLM4000系列,在由智能控制的消费电子和工业驱动等电路设计的众多领域中,成为全球工程师的首选工具。

示波记录仪

我们的示波记录仪是灵活、强大的多通道测试和测量解决方案,将高精度示波器和传统有纸图表记录仪的优点集于一身。Yokogawa的示波记录仪除了能在它的大显示屏上显示信号之外,还可以将信号保存到记录纸和/或存储介质中以便长期储存(如30天或更长),此外它还可以捕捉并分析非常快的瞬变信号。

高速数据采集器和数采仪

Yokogawa高速数据采集系统提供业界领先的隔离、位分辨率、采样速率和存储长度,带有独立的通道硬件和使用简单的软件。是示波记录仪中最领先的技术。

Precision Making

返回顶部
WeChat QR Code
微信扫一扫
获得更多专业服务