How do I Resolve the Warning Error 142 at Wavelength Calibration on AQ6319 Optical Spectrum Analyzer?

Warning Error 142 at Wavelength Calibration on AQ6319 Optical Spectrum Analyzer

The warning "142 WL calibration failed" is displayed when a level of light source is not enough at wavelength calibration or calibration cannot be executed because wavelength difference is out of calibration range.

There are three suggestions listed below.

  1. Please clean the connected fiber cable or exchange the cable for another
  2. Please initialize the alignment adjustment value and wavelength calibration value
  3. Please initialize the current parameter setting value, data, alignment adjustment value, and wavelength calibration value 

After restarting, please try the wavelength calibration again.

  1. Press [SYSTEM] and <PARAMETER INITIALIZE> key
  2. Select ALL CLEAR
  3. Select EXCECUTE
  4. Please confirm output level of standard light source (CW, around -22 dBm) by optical power meter

Produtos e Soluções Relacionadas

AQ6319 Optical Spectrum Analyzer

With high performance capabilities and a wavelength resolution of 10 picometer or less, the AQ6319 is designed for advanced research and development applications. With a wavelength range or 600 to 1700 nm this OSA offers an excellent combination of wavelength range and resolution.

Analisador de Espectro Ótico

Um analisador óptico de espectro (ou OSA) é um instrumento de precisão projetado para medir e exibir a distribuição de energia de uma fonte ótica em um intervalo de comprimento de onda específico. Um rastreamento OSA exibe a energia na escala vertical e o comprimento de onda na escala horizontal.

Equipamentos de teste ótico

A Yokogawa oferece soluções para medir componentes e sistemas óticos tradicionais e de ponta, atendendo à crescente demanda por linhas de fibra de alta capacidade e novas tecnologias de componentes.

Precision Making

Topo