When the Standard Operating Condition States "After 30 Minutes Warm-up and After Calibration," Does the "Calibration" Refer to the ScopeCorder's Auto-Cal Routine?

No, this "calibration" does not refer to the ScopeCorders Auto-Cal feature. When the Auto-Cal function is set to OFF, the ScopeCorder will not automatically calibrate after a 30 minute warm-up period. You must do a manual calibration after 30 minutes of warm-up. Press the Shift + Setup (Cal) keys to display the Cal menu. Then press the Execute Calibration soft key to execute a manual calibration.
 