Why Does the Sampling Rate Display Disappear During FFT Analysis for the DL1500 Digital Oscilloscope Series?

The FFT is computed based on the ACQ data. In order to perform the FFT over 1 k points, the ACQ data are decimated (reduced) from 10 kW to 1 kW. At this point the sampling rate is equivalent to 100 kSs (1/10 of 1 MS/s). Therefore, the sampling rate display is turned OFF. For example, to perform an FFT over the data sampled at 1 MS/s when the memory length is 10 kW, the data are decimated from 10 kW down to 1 kW. This means that the sampling rate is equivalent to 100 kS/s at this point. Therefore, to avoid misinterpretation, the sampling rate display is turned OFF. In this case, the actual sampling rate is 1 MS/s, which means that the frequency bandwidth of the time axis waveform is 500 kHz. However, because the sampling rate is reduced to 1/10 as explained above, the frequency bandwidth of the FFT is 50 kHz.

Produtos e Soluções Relacionadas

Mixed Signal Oscilloscopes

  • Yokogawa mixed signal oscilloscopes
  • Simultaneous, time correlated observations
  • Analysis of analog with digital (logic) signals
  • Troubleshooting electrical anomalies, measuring parametric values, monitoring signals

Osciloscópios

Osciloscópios digitais de classe mundial da Yokogawa: Os osciloscópios digitais têm amostragem de alta velocidade e uma ampla gama de larguras de banda que podem ser utilizadas para projeto e desenvolvimento de dispositivos eletrônicos. As ScopeCorders têm as vantagens de um osciloscópio digital e de um gravador de dados multicanal.