Why Won't the Numerical Value I Enter for Record Length Save for the WE7275/WE7271/WE7272 Digitizer Module?

The minimum record length which can be input is limited such that "record length x sampling interval" must be 5 milliseconds or more. When you set the record length to a value smaller than this it automatically changes to a 0 or an otherwise allowable minimum record length value. Setting the value to 0 indicates "Maximum Record Length/Number of Memory Partitions".

Related Products & Solutions

WE7000 PC-Based Measurement Instruments

One system, multiple instruments:  WE7000 satisfies demands for fast, reliable and precise data acquisition which uses a standard laptop or PC as its user interface. Input modules plug into an expandable measuring station. Via the optical interface, it is possible to galvanically isolate the laptop/PC.

Data Acquisition (DAQ)

Data acquisition (DAQ) is the measurement, recording, analyzing, and presentation of real world phenomena. It includes electrical measurements such as voltage, current, power as well as measurements through sensors and transducers including temperature, pressure, strain, vibration, and more.

High Speed Data Acquisition

Yokogawa high speed data acquisition systems deliver industry leading isolation, bit resolution, sampling rate, and memory depth, with independent channel hardware and easy to use software.