Why Won't the Numerical Value I Enter for Record Length Save for the WE7275/WE7271/WE7272 Digitizer Module?

The minimum record length which can be input is limited such that "record length x sampling interval" must be 5 milliseconds or more. When you set the record length to a value smaller than this it automatically changes to a 0 or an otherwise allowable minimum record length value. Setting the value to 0 indicates "Maximum Record Length/Number of Memory Partitions".

Related Products & Solutions

Data Acquisition (DAQ)

Scalable DAQ systems with industry-leading isolation, noise immunity, built-in conditioning, and real-time analysis, ensuring accurate, reliable measurements and faster decisions.

High Speed Data Acquisition

  • PC-based, streaming, local, or remote operation
  • 20+ modules, isolated and versatile inputs
  • Up to 200 MS/s or 640 ch
  • Used in aerospace, automotive, energy, and manufacturing industries
  •  

WE7000 PC-Based Measurement Instruments

One system, multiple instruments:  WE7000 satisfies demands for fast, reliable and precise data acquisition which uses a standard laptop or PC as its user interface. Input modules plug into an expandable measuring station. Via the optical interface, it is possible to galvanically isolate the laptop/PC.