Why Won't the Numerical Value I Enter for Record Length Save for the WE7275/WE7271/WE7272 Digitizer Module?

The minimum record length which can be input is limited such that "record length x sampling interval" must be 5 milliseconds or more. When you set the record length to a value smaller than this it automatically changes to a 0 or an otherwise allowable minimum record length value. Setting the value to 0 indicates "Maximum Record Length/Number of Memory Partitions".

Related Products & Solutions

WE7000 PC-Based Measurement Instruments

One system, multiple instruments:  WE7000 satisfies demands for fast, reliable and precise data acquisition which uses a standard laptop or PC as its user interface. Input modules plug into an expandable measuring station. Via the optical interface, it is possible to galvanically isolate the laptop/PC.