How do I Update the Voltage and Current Range on the WT100/WT200 Series Instrument Using IEEE488.2 Commands?

To change the voltage range on element 1 to the 30 V range, send the "RV4" command to the measuring instrument.
To change the current range on element 1 to the 1 A range, send the "RA5" command to the measuring instrument.

After entering settings, confirm the settings by sending the following commands:

  • To query the voltage: "RV?"
  • To query the current: "RA?"

For information on the correspondence between voltage values, current values, and parameters, see the following:

  • WT110E/WT130 Digital Power Meter User's Manual
    Page App1-8
  • WT200 Digital Power Meter User's Manual
    Pages 13-9 and 13-10
  • WT210/230M Digital Power Meter User's Manual
    Pages 13-11 and 13-12